Air Sampling Pumps

GilAir Plus Personal Air Sampling Pumps新葡新京产品

分享:

One Air Sampling Pump with All the Features You Need and Want

 3 New Technologies

  • Motion Detection
  • Bluetooth Connectivity
  • CONNECT Mobile App*

 Also Includes:

  • QuadModeSM Air Sampling Technology
  • Selectable Automatic Fault Recovery
  • Standard Temperature & Pressure Correction
  • Superior Design & Comfort
  • Real-time Clock with Data-logging

 

Innovation Delivered

The ultimate personal air sampling pump delivering enhanced performance, greater versatility, and superior design and comfort.

Motion Detection

Detect and measure up to 3 levels of activity (active, stationary, inactive) of each worker, ensuring sample integrity

Bluetooth Connectivity

  • Remotely monitor an air sampling pump up to 5 meters away
  • No disturbance of the worker
  • Stay in unrestricted areas while workers are in restricted areas 
  • No need to suit up

CONNECT Mobile App

  • Start, stop or pause the sample remotely
  • Monitor and graph flow rate, sample run time, and volume remotely
  • Perform real-time remote assessments:
    • Identify if cassette is overloading
    • Identify if filter media is detached
  • Sample audits provide the following right on the smart phone/tablet:
    • Report of current conditions
    • Screen shot of current conditions
    • Time stamp
    • GPS coordinates (actual address)
    • Pump number
    • E-mailable report
  • Check the status of multiple pumps in the same area. 

*List of tested iOS and Android Devices:

  Type

    OS Version   

  Tested Devices

 iOS

  8.0 to 10

  苹果s 5, 5C, 5S, 6, 6+, 6S, 6S+, iPod Touch and iPad

   Android  

4.2 to 6.0

  Samsung S3, S4, S5, Note 3, Moto E, Nexus and Samsung Tablets  

This list will be updated as new smart phone/tablet models are released and assessed.

Please contact Sensidyne ’s Software Support Team at office@gongle.net with questions regarding additional devices or support for this application.

 

产品图片

Product Specifications

Performance

Flow Range 
Constant flow 20 - 5,000 cc/min. without external adaptors 
Constant pressure 1 - 5,000 cc/min without external adaptors

Flow Modes 
Constant Flow or High and Low Flow Constant Pressure

Accuracy

Flow Display ± 5% or 3 cc/min of set flow, whichever is larger

Constant Flow Control ± 5% or 3 cc/min of set flow, whichever is larger

Constant Pressure Control ± 10% of back-pressure

Back Pressure Capability

(8-hour runtime) 5000 cc/min. up to 12” H20 back pressure

4000 cc/min. up to 20” H20 back pressure
3000 cc/min. up to 30” H20 back pressure
2000 cc/min. up to 30” H20 back pressure
1000 cc/min. up to 35” H20 back pressure
450 - 1000 cc/min. up to 40” H20 back pressure
20 - 449 cc/min. up to 40” H20 back pressure

Run Time 8 hour minimum

Flow Fault If flow changes exceed 5% within back pressure specifications, fault notification appears. If fault exceeds 30 seconds, pump shuts down. Selectable automatic fault recovery allows the pump to attempt restart every 3 minutes for up to 10 times or to hold until manual intervention

Environmental

Temperature Ranges

Operating 32°F to 113°F (0°C to 45°C)

Storage -4°F to 113°F (-20°C to 45°C)

Charging 41°F to 104°F (5°C to 40°C)

Humidity Ranges

Operating 5-95 %RH, non-condensing

Storage 5-98 %RH, non-condensing

GENERAL

Display Flow rate, sample time, and sample volume in actual conditions

Controls

Indicators Green and Red LED’s

Dimensions 4.3 x 2.4 x 2.4 inches (10.9 x 6.1 x 6.1 centimeters)

Weight 20.5 ounce (581 grams)

Warranty 2 years from original shipment, 5 years for pump keypad, and 1 year on rechargeable NiMH battery pack

ELECTRICAL

Power Options Removable rechargeable nickel-metal hydride (NiMH) battery pack, optional alkaline battery pack, or DC power

Battery Level Indicator On-screen icon displays battery level

Interface Connectors All pump interface is carried out via the Dock or keypad

Dock Functions Battery charging, computer interface, and auto-calibration

Charging Time Under 3.5 hours

APPROVALS

EN 1232 Compliant Ingress Protection: IP54

Intrinsic Safety

US/Canada NRTL (FM) Class I, Div 1 Groups A, B, C, D, E, F, and G;
Class II, Div 1 Zone 0; Class III, T4

ATEX/IECEx
 II 1 G
Ex ia IIC T4

CE

Applicable Directives ATEX 94/9/EC
EMC 2004/108/EC (RFI) LVD 2006/95/EC

RoHS Compliant